What is the working principle of the drug test box?
What is the working principle of the drug test box?
The box structure of the drug test box adopts a curved design, and the shape is novel and beautiful. The drug test box is suitable for pharmaceutical companies and pharmaceutical scientific research units to test and verify the stability of drugs and new drugs, so what is the working principle of the drug test box?

What is the working principle of the drug test box ?
1. Temperature system control mode:
Balanced temperature control methods with forced recirculation ventilation. The control system controls the output of the heater at the set temperature point according to the result of PID automatic calculation, so as to achieve dynamic balance. The humidity controller outputs signals to the control solid state relay through the internal PID integrated block electronic device, and then controls the power supply of the humidifier through the normally closed and normally open solid state relays. When the PT100 directly senses the SV set temperature, the humidifier stops humidifying and maintains a constant humidity state.
2. Air heating mode:
Stainless steel heater, good material, corrosion resistance, moisture resistance, speed response greatly exceeds the efficiency of the heating tube.
3. Humidification:
Stainless steel electric heating tubes are used to directly heat water to extract moisture, and an over-temperature protection system is added to the design. Humidification system: Tap water is automatically filtered through carbon and substance rod filters, and then the solenoid valve is controlled by the water level float switch to replenish water.
What are the preparations before the work of the drug test box?
1. Put the box on the dry ground and workbench indoors, do not use other fixing devices. After the power is turned on, the heating switch can be turned on, and the oven will set the temperature setting dial to the required working temperature value. Set the temperature "setting-measurement" switch in the constant temperature incubator and waterproof incubator to "setting", adjust the temperature setting knob to the required working temperature, and then set the "setting-measurement" switch to "measurement". At this time, the temperature in the incubator starts to rise, and the green light turns on.
2. Before power on, check the electrical performance of the box, and pay attention to whether there is an open circuit or leakage. After everything is ready, put the sample in, close the door, and unscrew the exhaust valve of the drying box of the stability test box if necessary, with a gap of about 10 mm.
3. A switch knife or an air switch should be installed in the power supply line dedicated to this box, and a thick wire twice the cross-sectional area of the power line should be used as the grounding wire. When the oven is at a constant temperature, the high-temperature switch should be turned off, and only one set of electric heaters should be left to work, so as not to affect the accuracy of the constant temperature due to excessive power. When the working temperature is lower than 150, only one heater is allowed to work, and the high temperature switch is turned off.
4. When the temperature rises to the desired working temperature, turn off the green light, then make fine adjustments until the green light turns on again. The place where the indicator light turns on and off alternately is the constant temperature point. At this time, you can turn the knob to the place where the green light is off and keep it unchanged. When the temperature in the working room is stable, you can slightly adjust the thermostat to improve the temperature control accuracy. This method can choose any working temperature.
5. Waterproof incubator, connect distilled water or deionized water to the water inlet, and add water until it overflows. After reaching the temperature, the stability test box can be kept at constant temperature for a certain period of time according to the test requirements. During this process, the temperature can be automatically controlled by a thermostat without manual management.